23
7.5. br - Baud rate
Baudrate of the serial interface
Possible parameters:
1200 bauds
2400 bauds
4800 bauds
9600 bauds
7.6. P - Parity
Parity bits of the serial interface
Possible parameters:
0 = no parity
1 = odd parity
2 = even parity
7.7. H - Handshake
Handshake of the serial interface
Possible parameters:
0 = XOn/XOff, software handshake
1 = RTS/CTS, hardware handshake
7.8. Sb - Stand-By
STAND-BY
External stand-by for emergency switch-off
Possible parameters:
0 = stand-by input is ignored
1 = stand-by input is active
(connector see page 30)
7.9. Pr - Programmer type
REG+E-PROG
For adapting the programmer input to the connected programmer.
Possible parameters:
0 = voltage input 0 to 10 V
1 = current input 0 to 20 mA
2 = current input 4 to 20 mA
The voltage and current input ranges are fixed and correspond to a
temperature range from -100 °C to +400 °C.
Press the setpoint key
and the setpoint value (in °C) appears
on the MULTI-DISPLAY (LED).