39
Error messages Description
-20 WARNING: CLEAN CONDENSOR
OR CHECK COOLING WATER
CIRCUIT OF REFRIGERATOR
Cooling of the condenser is affected. Clean air-
cooled condenser.
Check the flow rate and cooling water
temperature on water-cooled condenser.
-21 WARNING: COMPRESSOR STAGE 1
DOES NOT WORK
Compressor stage 1 does not work.
-22 WARNING: COMPRESSOR STAGE 2
DOES NOT WORK
Compressor stage 2 does not work.
-23 WARNING: HIGH TEMPERATURE
ON COMPRESSOR STAGE 1
Excess temperature on compressor stage 1.
-24 WARNING: HIGH TEMPERATURE
ON COMPRESSOR STAGE 2
Excess temperature on compressor stage 1.
-25 REFRIGERATOR WARNING Error in the cooling machine.
-26 WARNING: STAND-BY PLUG IS
MISSING
External stand-by contact is open.
Stand-by input - see pages 23 and 30.
-30 CONFIGURATION ERROR:
CONFIRM BY PRESSING <ENTER>
ON CIRCULATOR
The configuration of the circulator does not
conform to its present use.
Press enter to automatically perform a
single modification of the configuration.
(Then contact an authorized service station).
12. Maintaining the cooling performance
To maintain the full cooling performance, clean the condenser from time
to time.
Switch off the unit, disconnect mains power cable.
Hold the venting grid, pull out and remove.
Clean the ribbed condenser with a vacuum cleaner.
Replace the venting grid.
Switch on the unit.