14. Technical specifications
HD SD
Temperature selection digital
via keypad indication on MULTI-DISPLAY (LED)
remote control via personal computer indication on monitor
Temperature indication MULTI-DISPLAY (LED)
Resolution °C 0.01 from -9.99 °C to +99.99 °C
°C 0.1 from ≤ -10.0 °C to ≥ +100.0 °C
Absolute Temperature Calibration:
ATC1 °C ±3
ATC2 °C ±9
Temperature control PID
Working temperature sensor Pt100
Safety temperature sensor Pt100
Heater wattage W 2000 (at 230 V) 2000 (at 230 V)
W 1000 (at 115V) 1000 (at 115V)
Pressure pump:
pressure, max. psi at 0 liters 5 8.2
discharge, max. Lpm at 0 psi 20 24
Suction pump:
suction, max. psi at 0 liters 3.2 5.8
discharge, max. Lpm at 0 psi 14 16
Electrical connections:
Computer interface RS232/RS485
Programmer input -100 °C to 400 °C = 0 - 10 V or 0 - 20 mA or 4 - 20 mA
Temperature recorder output 0 - 10 V (0 V = -100 °C, 10 V = 400 °C)
0 - 20 mA (0 mA = -100 °C, 20 mA = 400 °C)
4 - 20 mA (4 mA = -100 °C, 20 mA = 400 °C)
Stand-by input
Alarm output 24 to 0 V DC / max. 25 mA
External measurement and control sensor - Pt100
Control connectors for solenoid valves or supplementary pump - 230 V or 115 V / max. 1.25 A
Mains power connection V/ Hz 230 / 50 230 / 50
(±10 %) V/ Hz or 115 / 60 or 115 / 60
Total power consumption (at 230 V) Watts 2400 2400
(at 115 V) Watts 1400 1400
All measurements have been carried out at:
rated voltage and frequency ambient temperature: 20 °C
Technical changes without prior notification reserved.