Remote control
34
11.3. List of commands
When the RS485 interface is used, the instrument address stands in front of each command (Axxx_).
in commands: Asking for parameters or temperature values to be displayed.
Command Parameter Response of circulator
version none Number of software version (V X.xx)
status none Status message, error message (see below 36).
in_pv_00 none Actual bath temperature.
in_pv_01 none Heater wattage being used (%).
in_pv_02 none Temperature value registered by the external Pt100 sensor.
in_pv_03 none Temperature value registered by the safety sensor.
in_sp_00 none Working temperature "T1"
in_sp_01 none Working temperature "T2"
in_sp_03 none
High temperature warning limit "
".
in_sp_04 none
Low temperature warning limit "
".
in_sp_05 none Setpoint temperature of the external programmer
(socket 21 - REG+E-PROG) .
in_hil_00 none Max. cooling capacity (%).
in_hil_01 none Max. heater capacity (%).
in_mode_01 none Selected working temperature:
0 = "T1".
1 = "T2".
in_mode_02 none Identification type:
0 = no identification
1 = single identification
2 = continual identification