7.10. Ad - Address
For determination of a circulator address for operation via the
serial RS485 interface.
(example: Ad32)
Possible parameters: 0 to 99
In case a superordinated data system controls several circulators at
one time, commands are directed to the corresponding instrument
using the respective circulator address.
The address must stand in front of each command.
7.11. Signal level RS232/RS485
The serial interfaces RS232 and RS485 use different output levels.
When the circulator is used in keypad control mode and values are
to be called off via the serial interface at the same time, use this
menu item for setting the desired signal level.
Possible parameters: c232
c485
If under the menu item „Remote“ (see page 22) one of the serial
interfaces is selected, the signal level is automatically set.