Command Parameter Response of circulator
out_par_04 xxx Bandwidth limit during external control.
Setting the maximum difference between the temperatures
in the internal bath and external system.
out_par_05 xxx Ratio between max. cooling capacity versus max. heater capacity
(0...0.99).
out_par_06 xxx Xp control parameter of the internal controller.
out_par_07 xxx Tn control parameter of the internal controller.
out_par_08 xxx Tv control parameter of the internal controller.
out_par_09 xxx Xp control parameter of the cascade controller.
out_par_10 xxx Proportional portion of the cascade controller.
out_par_11 xxx Tn control parameter of the cascade controller.
out_par_12 xxx Tv control parameter of the cascade controller.
11.4. Status messages / error messages
The circulator sends data (including error messages) only when the computer sends a query.
Status messages Description
00 MANUAL STOP Circulator in "OFF" state.
01 MANUAL START Circulator in keypad control mode.
02 REMOTE STOP Circulator in "r OFF" state.
03 REMOTE START Circulator in remote control mode.