Remote control
36
out commands: Setting parameters or temperature values.
Command Parameter Response of circulator
out_mode_01 0 Use working temperature "T1" *
out_mode_01 1 Use working temperature "T2" *
out_mode_02 0 No identification.
Temperature control by using the stored parameters.
out_mode_02 1 Single identification of controlled system after the next start.
out_mode_02 2 Continual identification of controlled system whenever a new
setpoint is to be reached.
out_mode_04 0 Temperature control of internal bath.
out_mode_04 1 External control with Pt100 sensor.
out_mode_05 0 Stop the circulator = r OFF.
out_mode_05 1 Start the circulator.
out_sp_00 xxx.x Set working temperature "T1".
out_sp_01 xxx.x Set working temperature "T2".
out_sp_03 xxx.x
Set high temperature warning limit .
out_sp_04 xxx.x
Set low temperature warning limit .
out_hil_01 xxx Set the desired maximum heater capacity (10 % to 100 %)
(* see "Note" on page 20)