35
Command Parameter Response of circulator
in_mode_03 none Type of the programmer input:
0 = Voltage 0 V to 10 V
1 = Current 0 mA to 20 mA
2 = Current 4 mA to 20 mA
in_mode_04 none Internal/external temperature control:
0 = Temperature control in the circulator bath.
1 = Temperature control with external Pt100 sensor.
in_mode_05 none Circulator in Stop/Start condition:
0 = Stop
1 = Start
in_par_01 none Time constant of the external bath.
in_par_02 none Internal slope.
in_par_03 none Time constant of the internal bath.
in_par_04 none Bandwidth limit (max. difference between the temperatures in the
internal bath and external system).
in_par_05 none Ratio for max. cooling capacity versus max. heating capacity.
in_par_06 none Xp control parameter of the internal controller.
in_par_07 none Tn control parameter of the internal controller.
in_par_08 none Tv control parameter of the internal controller.
in_par_09 none Xp control parameter of the cascade controller.
in_par_10 none Proportional portion of the cascade controller.
in_par_11 none Tn control parameter of the cascade controller.
in_par_12 none Tv control parameter of the cascade controller.