ME
17
JULABO
Description
Thermal
H10
Thermal
H20S
Thermal
H200
Order Number 10 liters 8 940 114 8 940 108 8 940 134
5 liters 8 940 115 8 940 109 8 940 135
Temperature range °C -20 ... 180 0 ... 220 60 ...200
Flash point °C 190 230 292
Fire point °C 216 274 334
Color clear light brown clear
See website for list of recommended bath fluids.
ATTENTION: The maximum permissible viscosity is 50 mm
2
/s
.
Caution:
Fire or other dangers when using bath fluids that are not recommended:
Please contact JULABO before using other than recommended bath fluids.
Use only nonacidic and noncorrosive bath fluids.
JULABO assumes no liability for damage caused by the selection of an unsuitable
bath liquid.
Unsuitable bath fluids are fluids which, e.g.,
• are highly viscous
(much higher than recommended at the respective working temperature)
• have a low viscosity and have creep characteristics
• have corrosive characteristics or
• tend to crack.
• No liability for use of other bath fluids!
5.3. Temperature application to external systems
Caution: Securely attach all tubing to prevent slipping.
If the circulator is operated without external system, close the pump connector (24a)
with the cap nut.
feed return
The circulator is used for temperature application to external,
closed systems (loop circuit) with simultaneous temperature
application in the circulator bath.
Connecting the external system
• Unscrew the collar nuts from the pump connector (24a).
• Slide the tubing onto the pump connector for feed (24a) and
return flow (24b) and secure with hose clamps.