ME
29
9.1. MENU PROGRAM – START
Start-Menu
This menu will start a previously set program.
L Requirements:
1. Create a program. (refer to next chapter)
2. Return to the Start-MENU and confirm the desired setting of each
MENU item with the key
3. Set a start time (>TIME< >DATE< >YEAR<) if the program is to be
started by the internal timer.
Menu level 1
> STEP< Program start at section 1 … 10
> RUNS < Number of repetitions 1 ... 99
> END< Status at end of program (STDBY/SETPNT)
Standby or last setpoint
> GO < Time of start (NOW/TIMER)
Level 2 Parameter level
L Press the key if a parameter is to be retained.
Correction function for parameters or values (prior to
OK)
Set program step with
and
example: STEP 1
Set number of runs with
and
example: 1 run
or
Set desired parameters with
and .
L (STDBY / SETPNT)
Parameter STanDBY: the circulator switches to –
OFF-.
Parameter SETPoiNT: the circulator constantly
keeps the temperature at the value of the last step.
or
É
Confirm >NOW< with the key and the program
will start immediately
L or
start at the set time under parameter (TIMER ).
Set time in the example below:
09. August 2009, 11:15 hrs