EasyManua.ls Logo

Julabo ME - 9.7.3. CALIBRATION TYPE: 1 - 2 -; 3 POINT

Julabo ME
62 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ME
43
Menu level 1
Level 2 Parameter level
L Press the key if parameter is to be retained.
Correction function for parameters or values (prior
to OK).
or
The parameter flashes, switch by pressing
and
L On level 2 a (I) is indicated for internal or an (E)
for external.
Example:
or
The parameter flashes, switch by pressing
and
L >NO< Carry out an ATC calibration
L >YES< return to standard operation after
calibration.
The parameter flashes, switch by pressing
and
L A >1-point<, >2-point< or >3-point< calibration
can be carried out.
The selected calibration is indicated on level 2 by 1 or
2 or 3.
The value >TMPVAL< is only indicated
L In addition the measured temperature value
>CALVAL X< is saved during the next step.
Integer digits flash, set by pressing
+
Decimal digits flash, set by pressing
+
L If only a 1-point calibration is carried out, the following menu items are not
indicated anymore

Table of Contents

Related product manuals