Menu functions
50
BAND HIGH / LOW – Band limitation
The band limitation is active during external control. Varied, practice-
oriented setting are feasible for heat-up and cool-down phases.
Setting range:
0 °C ... 200 °C
BAND HIGH and BAND LOW allow for the limitation of the difference
between the temperatures in the internal bath and the external system to
any maximum value for the heat-up and cool-down phase.
During the heat-up phase this difference value is always added to the
actual external temperature. During the cool-down phase the difference
value is subtracted.
25
50
10 20 30 40 50 60 10 20 30 40 50 60 10
[
min
]
75
100
150
200
°C
t
System extern
Bad intern
BandOben
BandUnten
Sense of a band limitation:
Protection of objects and samples by
gentle temperature control
Protection of e.g. glass reactors from
thermal shock.