4 Technical data
24
Absolute pressure
Nominal measuring
range
0 to 600 mbar 0 to 2.5 bar 0 to 4 bar 0 to 10 bar
Default
measuring range
0 to 600 mbar 0 to 2.5 bar 0 to 4 bar 0 to 10 bar
Smallest MSP
a
60 mbar 0.1 bar 0.1 bar 0.5 bar
Turndown ratio (r)
b
r ≤ 10 r ≤ 20 r ≤ 40 r ≤ 20
Non-linearity for refer-
ence conditions
c
0.12 % 0.04 % 0.04 % 0.04 %
Accuracy in %
of the set MSP
Range: -20 to +60 °C
r × 0.24 %
for 1 ≤ r ≤ 10
r×0.08%
for 1 ≤ r ≤ 20
r×0.08%
for 1 ≤ r ≤ 40
r × 0.08 %
for 1 ≤ r ≤ 20
Accuracy in %
of the set MSP
Range: -40 to -20 °C
or 60 to 85 °C
r × 0.36 %
for 1 ≤ r ≤ 10
d
r×0.16%
for 1 ≤ r ≤ 20
r×0.16%
for 1 ≤ r ≤ 40
r × 0.16 %
for 1 ≤ r ≤ 20
Long-term stability in %
as of the nominal mea-
suring range
0.1 %/year
Absolute pressure
Nominal measuring
range
0 to 25 bar 0 to 100 bar
Default
measuring range
0 to 25 bar 0 to 100 bar
Smallest MSP
a
0.5 bar 5 bar
Turndown ratio (r)
b
r ≤ 50 r ≤ 20
Non-linearity for refer-
ence conditions
c
0.04 % 0.04 %
Accuracy in %
of the set MSP
Range: -20 to +60 °C
r × 0.08 %
for 1 ≤ r ≤ 50
r×0.08%
for 1 ≤ r ≤ 20
Accuracy in %
of the set MSP
Range: -40 to -20 °C
or 60to8C
r × 0.16 %
for 1 ≤ r ≤ 50
r×0.12%
for 1 ≤ r ≤ 20
Long-term stability in %
as of the nominal mea-
suring range
0.1 %/year
a
MSP = measuring span
b
r = span of the nominal measuring range ÷ adjusted measuring span
c
Reference conditions: ambient temperature 20 °C (±3 K), air pressure 1000 hPa (±25 hPa)
d
Only up to -30 °C