2 General information
2.2 Scope of delivery
Operating Manual
The operating manual describe the mounting, the electrical connection, the startup and operation of the
device.
Safety manual (option)
The safety manual describes the safe application with the installation according to IEC 61508.
Calibration certificate
The device is supplied with a calibration certificate and a setup print-out.
These documents contain information about the set parameters and/or the measured parameters for the
respective device.
If the calibration certificate is lost or if you need another copy, the calibration certificate can be requested
from the manufacturer by specifying the device's F number (see nameplate). You will find the supplier's
address on the back cover of the manual.
Setup program (option)
The setup program is available as an option: part no. 00537577
All the device's parameters can be conveniently checked and adjusted using the setup program - there
are also additional functions, e.g.:
Recording the measured values
Graphical view of temperature and pressure
Detailed diagnostic messages
Display of the complete order code and the device configuration (for follow-up orders)
The setup program accesses the device via
the JUMO interface (standard) or
the HART® interface (optional)
DANGER!
The JUMO interface must not be used in the Ex-area!
The device may only be operated using the rotary knob or the HART® interface!
PC interface cable (option)
Available as an accessory: PC interface cable with USB/TTL converter and two adapters (USB transmit-
ter cable), part no. 00456352.
The device can be connected to a PC's USB port via the JUMO interface with the PC interface cable.
HART® modem (option)
Available as an accessory: HAR modem for USB, part no. 00443447.
The device can be connected to a PC's USB® port via the HART® interface with the HART® modem.
Input isolating amplifier (option)
Available as an accessory: Input isolating amplifier for Ex-applications, HART®-capable, part no.
00577948.
Devices with explosion protection ATEX Ex ia must be connected via an input isolating amplifier for use
in the Ex-area!