Calibration Command Reference
3-11
:CALCulate
(:CALibration:PROTected:DC):CALCulate
Purpose To program the Model 2001 to calculate new comprehensive calibration DC constants.
Format :cal:prot:dc:calc
Parameters None
Description The :CALCulate command should be sent to the instrument after performing all other DC
calibration steps to calculate new comprehensive calibration constants. All other compre-
hensive calibration steps must be completed before sending this command.
Programming example 10 OUTPUT 716; “:CAL:PROT:DC:CALC” ! Calculate new constants.