EasyManuals Logo
Home>Keithley>Multimeter>2001

Keithley 2001 Calibration Manual

Keithley 2001
107 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
Calibration Command Reference
3-11
:CALCulate
(:CALibration:PROTected:DC):CALCulate
Purpose To program the Model 2001 to calculate new comprehensive calibration DC constants.
Format :cal:prot:dc:calc
Parameters None
Description The :CALCulate command should be sent to the instrument after performing all other DC
calibration steps to calculate new comprehensive calibration constants. All other compre-
hensive calibration steps must be completed before sending this command.
Programming example 10 OUTPUT 716; “:CAL:PROT:DC:CALC” ! Calculate new constants.

Table of Contents

Other manuals for Keithley 2001

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2001 and is the answer not in the manual?

Keithley 2001 Specifications

General IconGeneral
BrandKeithley
Model2001
CategoryMultimeter
LanguageEnglish

Related product manuals