Calibration Command Reference
3-12
3.4 :CALibration:UNPRotected Subsystem
3.3.11 :ACCompensation
(:CALibration:UNPRotected):ACCompensation
Purpose To perform user AC calibration.
Format :cal:unpr:acc
Parameters None
Description The :ACC command performs user AC calibration, which requires no calibration equip-
ment. All test leads must be disconnected from the input jacks when performing user AC
calibration.
Programming note Calibration constants generated by using the :ACC command are not stored in EEPROM.
Thus, AC calibration constants are in effect only until the instrument is turned off. In order
to save AC calibration constants, perform the comprehensive calibration procedure, and use
the :SAVE command.
Programming example 10 OUTPUT 716; “:CAL:UNPR:ACC: ! Perform AC user cal.