Calibration Command Reference
3-5
3.3.5 :DATE
(:CALibration:PROTected):DATE
Purpose To send the calibration date to the instrument.
Format :cal:prot:date “<string>”?
Parameters <string> = date (mm/dd/yy)
Description The :DATE command allows you to store the calibration date in instrument memory for fu-
ture reference. You can read back the date from the instrument over the bus by using the
:DATE? query, or by using the CALIBRATION selection in the front panel menu.
Programming note The date <string> must be enclosed either in double or single quotes (“<string>” or
‘<string>’).
Programming example 10 OUTPUT 716; “:CAL:PROT:DATE ‘01/01/92’”! Send date.
3.3.6 :DATE?
(:CALibration:PROTected):DATE?
Purpose To request the calibration date from the instrument.
Format :cal:prot:date?
Response <date> (mm/dd/yy)
Description The :DATE? query allows you to request from the instrument the previously stored calibra-
tion date. The instrument response is simply a string of ASCII characters representing the
last stored date.
Programming example 10 OUTPUT 716; “:CAL:PROT:DATE?” ! Query for date.
20 ENTER 716; A$ ! Input date.
30 PRINT A$ ! Display date.
3.3.7 :NDUE
(:CALibration:PROTected):NDUE
Purpose To send the next calibration due date to the instrument.
Format :cal:prot:ndue “<string>”
Parameters <string> = next due date (mm/dd/yy)
Description The :NDUE command allows you to store the date when calibration is next due in instru-
ment memory. You can read back the next due date from the instrument over the bus by us-
ing the :NDUE? query, or by using the CALIBRATION-DATES selection in the front panel
menu.
Programming note The next due date <string> must be enclosed either in single or double quotes (“<string>”
or ‘<string>’).
Programming example 10 OUTPUT 716; “:CAL:PROT:NDUE ‘01/01/93’” ! Send due date.

Table of Contents

Other manuals for Keithley 2001

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2001 and is the answer not in the manual?

Keithley 2001 Specifications

General IconGeneral
Model2001
CategoryMultimeter
ManufacturerKeithley
TypeDigital Multimeter
Digits of Resolution7.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
ConnectivityGPIB, RS-232
DisplayVFD
Resistance Accuracy0.008% + 2.0 mΩ
Frequency Accuracy±0.01%
DC Current Range10 μA to 3 A
AC Current Range10 μA to 3 A
Power Supply100 V to 240 V AC, 45 Hz to 66 Hz

Related product manuals