EasyManuals Logo
Home>Keithley>Multimeter>2002

Keithley 2002 Repair Manual

Keithley 2002
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #35 background imageLoading...
Page #35 background image
Troubleshooting
2-22
Test 105.8 — Group Execute Trigger (GET)
Type
Pass/Fail
Fault message
Open IEEE-488 GET
Description
Trigger 8 (STI8) is set up as an input, and the 9914 (U622) is then programmed to generate a
GET signal.
Drawing reference
Digital Board; 2002-140
Components
U618 and U622.
Test 105.11 through 105.18 — Trigger shorts
Type
Pass/Fail
Fault message
None
Description
In test 105.11, STI1 is programmed as the trigger input. Each of the other triggers (STI2 through
STI7 and GET) is programmed to toggle in sequence. If a trigger is detected at STI1, a short is
indicated and the test fails. The test is repeated for STI2 through STI8 in tests 105.12 through
105.18.
Ordinarily, Built-In-Tests should be run with no external connections. However, the Trigger
Shorts tests may be used to verify proper operation at the external trigger bus connector (J1029,
J1030). For example, by shorting pins 1 and 2, tests 105.11 and 105.12 should fail, indicating
that the short was detected between system trigger 1 and system trigger 2.
Drawing reference
Digital Board; 2002-140
Components
J1029, J1030, and associated circuitry.

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
BrandKeithley
Model2002
CategoryMultimeter
LanguageEnglish

Related product manuals