2-4 Network Connection Model 2701 Instrument Networking Instruction Manual
Hub connection
This section contains information on connecting the PC to one or more Model 2701s
through an Ethernet hub (
Figure 2-2) or switch that is not part of a larger local area network
using Ethernet straight through cables. When using this connection scheme, the TCP/IP set-
tings for the PCs Ethernet card and the Model 2701 will need to be set manually (see
Section 3).
NOTE Different available IP addresses are required for each Model 2701 and the network
interface card connected to the Model 2701 network.
Figure 2-2
Hub to Model 2701
Table 2-3
Additional component requirements for hub connection
Quantity Component Note
1 Network Interface Card Properly installed in PC.
2 CAT 5 straight-thru cable Add one additional cable for each additional
Model 2701 connected through the hub.
1 Network Hub 10/100BaseTX
NOTE:
The maximum distance between
the PC and hub and between
the hub and 2701(s) is 100m.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm

Related product manuals