IEEE-488 Reference
3-43
:CALCulate3
:LIMit[1]
:UPPer
[:DATA] <n>
[:DATA]?
:SOURce <NRf>
:SOURce?
:LOWer
[:DATA] <n>
[:DATA]?
:SOURce <NRf>
:SOURce?
:STATe <b>
:STATe?
:FAIL?
:CLEAR
[:IMMediate]
:AUTO <b>
:AUTO?
:LIMit2
:UPPer
[:DATA] <n>
[:DATA]?
:SOURce <NRf>
:SOURce?
:LOWer
[:DATA] <n>
[:DATA]?
:SOURce <NRf>
:SOURce?
:STATe <b>
:STATe?
:FAIL?
:CLEAR
[:IMMediate]
:AUTO <b>
:AUTO?
:PASS
:SOURce <NRf>
:CLIMits
:FAIL?
:BSTRobe
:STATe <b>
:STATe?
:IMMediate
Subsystem to control CALC 3 (limit tests):
Path to control LIMIT 1 test:
Path to configure upper limit:
Set upper limit (-9.999999e35 to +9.999999e35).
Query upper limit.
Specify digital output (0 to 15).
Query source.
Path to configure lower limit:
Set lower limit (-9.999999e35 to +9.999999e35).
Query lower limit.
Specify digital output (0 to 15).
Query source.
Enable or disable limit test.
Query state of limit test.
Query test result (1 = pass, 0 = fail).
Path to clear failed test:
Clear failed test indication.
Enable or disable auto clear.
Query auto clear.
Path to control LIMIT 2 test:
Path to configure upper limit:
Set upper limit (-9.999999e35 to +9.999999e35).
Query upper limit.
Set digital output (0 to 15).
Query source (digital output value).
Path to configure lower limit:
Specify lower limit (-9.999999e35 to +9.999999e35).
Query lower limit.
Specify digital output (0 to 15).
Query source (digital output value).
Enable or disable limit test.
Query state of limit test.
Query test result (1 = pass, 0 = fail).
Path to clear failed test:
Clear failed test indication.
Enable or disable auto clear.
Query auto clear.
Path to define “pass” digital output pattern.
Specify digital output (0 to 15).
Command path for composite limits:
Query composite result of LIMIT 1 and LIMIT 2 tests
(1 = pass, 0 = fail).
Path to control limit binning strobe:
Enable or disable strobe.
Query state of binning strobe.
Re-perform limit tests.
1
0
-1
0
OFF
ON
1
0
-1
0
OFF
ON
0
OFF
3.13.3
Note: ON is the *RST default parameter, and OFF is the :SYSTem:PRESet default.
Table 3-4 (Continued)
CALCulate command summary
Command Description
Default
parameter SCPI Ref.

Table of Contents

Other manuals for Keithley 6517A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517A and is the answer not in the manual?

Keithley 6517A Specifications

General IconGeneral
BrandKeithley
Model6517A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals