EasyManua.ls Logo

Keithley 6517A - Page 85

Keithley 6517A
343 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Front Panel Operation
2-64
Standard Method Resistivity tests (Surface and Volume)
This test is used to measure the resistivity (surface or vol-
ume) of an insulator sample. When used with the Model
8009 Resistivity Test Fixture, the test conforms to the ASTM
D-257 standard. For detailed information on resistivity mea-
surements, refer to paragraph 2.7.2. Figures 2-33 and 2-35
show the test circuits for the respective measurement, and
Figure 2-36 shows the connections to the Model 8009. Refer
to the instruction manual for the Model 8009 to install the in-
sulator sample in the test fixture.
When this test is run, the V-Source will initially be set to
source 0V for a specified time (PRE-DISCH time) to allow
any charge to dissipate. The V-Source will then apply a spec-
ified voltage (BIAS V) to the electrodes of the test fixture for
a specified time (BIAS-TIME). This “bias” period allows
currents in the test circuit to stabilize. The V-Source then ap-
plies the test voltage (MEAS-V) and, after a specified delay
(MEAS-TIME), the Model 6517A measures the resistivity
of the sample and stores the reading in the buffer. Note that
the test voltage (MEAS-V) is typically at the same level as
the bias voltage (BIAS V).
HI
LO
6517A
Shield
Resistor
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
7078-TRX
Cable
INPUT
250V PEAK
COMMON
TRIGGER
LINK
IN OUT
!
LINE RATING
90-134VAC
180-250VAC
50, 60, 400HZ
55VA MAX
LINE FUSE
SLOWBLOW
1/2A, 250V
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
PREAMP OUT
250V PEAK
V SOURCE
LO HI
Note: Ammeter LO internally
connected to V-source LO
(see paragraph 2.9.1).
A) Connections
6517A
V-Source
A
+
-
6517A
Ammeter
B) Equivalent Circuit
HI
LO
HI
LO
Resistor
DUT
Shield
Figure 2-51
Test circuit; resistor voltage coefficient test
The Surface Resistivity Test and the Volume Resistivity Test
are selected and configured from the CONFIGURE SE-
QUENCE menu (R/RESISTIVITY; NORMAL; SURFACE
and VOLUME). See paragraph 2.14.2 for details.
Alternating Polarity Resistance/Resistivity test
The Alternating Polarity Resistance/Resistivity test is de-
signed to improve high resistance/resistivity measurements.
These measurements are prone to large errors due to back-
ground currents. By using an alternating stimulus voltage, it
is possible to eliminate the effects of these background cur-
rents. This test will measure Surface or Volume resistivity, or
Resistance, as selected in the CONFIGURE RESISTANCE
menu. For detailed information on resistivity measurements,
refer to paragraph 2.7.2. Figures 2-33 and 2-35 show the test
circuits for the respective measurements, and Figure 2-36
shows the connections to the Model 8009. Refer to the Mod-
el 8009 Instruction Manual for information on installing the
sample in the test fixture.
When this test is run, the V-Source will alternate between two
voltages (V-OFS + V-ALT) and (V-OFS - V-ALT) at timed inter-

Table of Contents

Other manuals for Keithley 6517A

Related product manuals