2400 Series SourceMeter
®
User’s Manual Range, Digits, Speed, and Filters 7-7
Speed
The Speed/Accuracy menu is used to set the integration time of the A/D converter (period
of time the input signal is measured). The integration time affects the usable digits, the
amount of reading noise, and the ultimate reading rate of the instrument. The integration
time is specified in parameters based on the number of power line cycles (NPLC), where 1
PLC for 60Hz is 16.67msec (1/60) and 1 PLC for 50Hz and 400Hz is 20msec (1/50).
In general, the fastest integration time (FAST; 0.01 PLC) results in increased reading noise
and fewer usable digits. The slowest integration time (HI ACCURACY; 10 PLC) provides
the best noise rejection. In-between settings are a compromise between speed and noise.
The default power-on speed setting is NORMAL (1 PLC).
NOTE For the Pulse Mode of the Model 2430, the valid NPLC range is 0.01 to 0.1
PLC.
The speed setting affects both the NMRR (normal mode rejection ratio) and
CMRR (common mode rejection ratio). Normal mode noise is the noise signal
between the HI and LO input terminals, while common mode noise is the noise
signal between LO and chassis ground. See Appendix A for NMRR and CMRR
specifications.
Setting speed
Speed is set from the Speed Accuracy menu and is structured as follows. Use Section 1,
Rules to navigate menus,” to check and/or change the speed setting.
NOTE For the Pulse Mode of the Model 2430, speed is set from Pulse Speed menu. This
menu structure is located after the Speed Accuracy menu. Use the “Rules to nav-
igate menus” in Section 1 to check and/or change the speed setting.
Front panel speed control
Press SPEED or CONFIG SPEED to display the menu. (See Figure 7-1.)
•FAST — Sets speed to 0.01 PLC and sets display resolution to 3½ digits.
MED — Sets speed to 0.10 PLC and sets display resolution to 4½ digits.
NORMAL — Sets speed to 1.00 PLC and sets display resolution to 5½ digits.
HI ACCURACY — Sets speed to 10.00 PLC and sets display resolution to 6½
digits.
•OTHER — Use to set speed to any PLC value from 0.01 to 10. Display resolution
is not changed when speed is set with this option.
NOTE After setting speed, display resolution can be changed using the DIGITS key.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2400 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2400 Series and is the answer not in the manual?

Keithley SourceMeter 2400 Series Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2400 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals