2400 Series SourceMeter
®
User’s Manual SCPI Command Reference 18-55
Example reading string
The example ASCII reading string shown in Figure 18-1 shows a mea-
surement of a 10k resistor, with the SourceMeter configured to Source
I Measure V. The voltage reading is the voltage measurement
(1.000236V), the current reading is the current source value (100, and
the operation was performed 72.826 seconds after the SourceMeter was
turned on (or after timestamp was reset). The status reading of 48,132
indicates that bits 2, 10, 11, 12, 13, and 15 of the status word are set.
SOURce2 <name>
:FORMat:SOURce2 <name> Set SOUR2 and TTL response formats
Parameters <name> = ASCii ASCII format
HEXadecimal Hexadecimal format
OCTal Octal format
BINary Binary format
Query :SOURce2? Query response format
Description This command controls the response format for all
CALC2:XXXX:SOUR2 and SOUR2:TTL queries in a manner similar
to formats set by the FORM:SREG command. See “Calculate sub-
systems,” page 18-24, and “SOURce subsystem,” page 18-72, for
details.
CALC data elements
CALCulate <item list>
:FORMat:ELEMents:CALCulate <item list> Set CALC data elements
Parameters <item list> = CALC Include CALC1 or CALC2 data
TIME Include timestamp
STATus Include status information
NOTE Each item in the list must be separated by a comma (for example, :CALCulate
CALC,TIME,STAT).
Query :CALCulate? Query CALC data element list
Description This command allows you to retrieve timestamp and status information
with the CALC1:DATA? And CALC2:DATA? queries. It also allows
you to retrieve timestamp and status information when TRACe:FEED is
set to CALC1 or CALC2. See “Calculate subsystems,” page 18-24, for a
complete description of CALC1 and CALC2. See “Data elements,
page 18-51, for a description of TIME and STATus information.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2400 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2400 Series and is the answer not in the manual?

Keithley SourceMeter 2400 Series Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2400 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals