2400 Series SourceMeter
®
User’s Manual SCPI Command Reference 18-63
RESistance:MODE <name>
[:SENSe[1]]:RESistance:MODE <name> Select ohms measurement mode
Parameters <name> = MANual Manual ohms mode
AUTO Auto ohms mode
Query :MODE? Query ohms mode
Description This command is used to select the ohms measurement mode. With
MANual ohms selected, the user must configure the source and measure
aspects of the operation. When the ohms function is selected, the ohms
reading is simply the result of the V/I calculation.
Range changes cannot be made in manual ohms.
With AUTO ohms selected, the SourceMeter will be configured to
Source I Measure V when the ohms function is selected. The current
source value and voltage measurement range used depends on the ohms
measurement range that is selected.
See Section 4, “Ohms Measurements,” for details on manual and auto
ohms.
RESistance:OCOMpensated <b>
[:SENSe[1]]:RESistance:OCOMpensated <b> Control offset-compensated ohms
Parameters <b> = 1 or ON Enable offset compensation
0 or OFF Disable offset compensation
Query :OCOMpensated? Query state of offset compensation
Description This command is used to enable or disable offset-compensated ohms.
When using the auto ohms measurement mode, the current source level
is automatically set. When using the manual ohms measurement mode,
you must set the source (V or I) output level.
See Section 4, “Offset-compensated ohms,” for details on making off-
set-compensated ohms measurements.
NOTE Offset-compensated ohms will disable when the :MEASure? command (for the
resistance function) or the :CONFigure:RESistance command is sent.
NOTE For the Model 2430 Pulse Mode, offset-compensated ohms cannot be enabled
using this command. Sending this command results in error +831.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2400 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2400 Series and is the answer not in the manual?

Keithley SourceMeter 2400 Series Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2400 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals