1-16 Getting Started 2400 Series SourceMeter
®
User’s Manual
Front panel tests
Use the TEST/FRONT-PANEL-TESTS selection of the main MENU to test various
aspects of the front panel. Test selections include:
KEYS — Front panel keys are tested. Pressing a key displays a message that iden-
tifies that key. Pressing EXIT twice cancels this test.
DISPLAY PATTERNS — Use this selection to turn on all display pixels and
annunciators. Subsequent key-presses cycle through tests that turn off annunciators
and corner pixels of each digit, turn on the rows of the top-left display digit, and
turn on all annunciators and pixels of each digit in a sequential manner. Press EXIT
to cancel this test.
CHAR SET — This test displays special characters. Press EXIT to cancel the test.
See “Menus,” page 1-20, for more information.
Default settings
By using appropriate menu selections, you can save and recall various instrument setups,
define the power-on configuration, or restore factory defaults as outlined below.
Saving and restoring user setups
You can save and restore up to five of your own user setups as covered below. This feature
provides a convenient way to save specific instrument configurations and then recall them
as needed. Note that you can also set up the SourceMeter to restore a specific user setup at
power-on. See “Power-on configuration,” page 1-17.
NOTE There are two types of setups. Instrument configuration (user) setups (covered
here) and source memory sweep setups (Section 10).
Saving setups
1. Select the various instrument operating modes you wish to save.
2. Press the MENU key, select SAVESETUP, then press ENTER.
3. From the SAVESETUP menu, select GLOBAL, then press ENTER.
4. From the GLOBAL SETUP MENU, select SAVE, then press ENTER.
5. Select the setup position (0-4) to save, then press ENTER to complete the process.
Restoring setups
1. Press the MENU key, select SAVESETUP, then press ENTER.
2. From the SAVESETUP menu, select GLOBAL, then press ENTER.
3. From the GLOBAL SETUP MENU, select RESTORE, then press ENTER.
4. Select the setup position (0-4) to restore, then press ENTER to complete the
process.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2400 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2400 Series and is the answer not in the manual?

Keithley SourceMeter 2400 Series Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2400 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals