10-28 Sweep Operation 2400 Series SourceMeter
®
User’s Manual
Call OutputCmd(intGPIB, “:CALC2:LIM1:STAT ON”) 'Set Limit1 on
Call OutputCmd(intGPIB, “:CALC2:LIM1:COMP:FAIL IN”)'Set Fail Mode to In
Compliance
Call OutputCmd(intGPIB, “:CALC2:LIM1:COMP:SOUR2 8”)'Set Digital Output
Pattern for Compliance
Failure
Call OutputCmd(intGPIB, “:CALC2:LIM2:STAT ON”) 'Set Limit2 on
Call OutputCmd(intGPIB, “:CALC2:LIM2:UPP 2E-2”) 'Set Upper Limit to 20mA
Call OutputCmd(intGPIB, “:CALC2:LIM2:LOW 1E-3”) 'Set Lower Limit to 10mA
Call OutputCmd(intGPIB, “:CALC2:LIM2:UPP:SOUR2 2”) 'Set Digital Output
Pattern for Upper Limit
#2 Failure
Call OutputCmd(intGPIB, “:CALC2:LIM2:LOW:SOUR2 3”) 'Set Digital Output
Pattern for Lower Limit
#2 Failure
Call OutputCmd(intGPIB, “:CALC2:CLIM:PASS:SOUR2 13”)'Set Digital Output
Pattern for Limit #2
Passing
Call OutputCmd(intGPIB, “:CALC2:CLIM:MODE GRAD”) 'Set Limit Results to
Grading
Call OutputCmd(intGPIB, “:CALC2:CLIM:BCON END”) 'Set Binning Control
to End
Call OutputCmd(intGPIB, “:CALC2:CLIM:PASS:SML NEXT”) 'Set SML Pass Location
Call OutputCmd(intGPIB, “:CALC2:CLIM:FAIL:SML 100”) 'Set SML Fail Location
Call OutputCmd(intGPIB, “:SOUR:MEM:SAVE 1”) 'Save in Source Memory
Location 1.
'Setup Source Memory Location 2
'------------------------------
Call OutputCmd(intGPIB, “*RST”) 'Restore GPIB default
conditions.
Call OutputCmd(intGPIB, “:SOUR:FUNC VOLT”) 'Current Source
Function.
Call OutputCmd(intGPIB, “:SENS:FUNC ‘CURR:DC’”) 'Current Sense
Function.
Call OutputCmd(intGPIB, “:SENS:CURR:PROT .105”) 'Set 105mA Compliance
Call OutputCmd(intGPIB, “:SENS:CURR:RANGE .1”) 'Set 100mA Current
Measure Range
Call OutputCmd(intGPIB, “:SOUR:DEL 1”) 'Set Source Delay to 1
Call OutputCmd(intGPIB, “:SOUR:VOLT 10”) '10V Source Voltage.
Call OutputCmd(intGPIB, “:CALC2:FEED CURR”) 'Send Current(A)
Readings to Buffer
Call OutputCmd(intGPIB, “:CALC2:LIM1:STAT ON”) 'Set Limit 1 on
Call OutputCmd(intGPIB, “:CALC2:LIM1:COMP:FAIL IN”) 'Set Fail Mode to In
Compliance
Call OutputCmd(intGPIB, “:CALC2:LIM1:COMP:SOUR2 5”) 'Set Digital Output
Pattern for Compli-
ance Failure
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2400 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2400 Series and is the answer not in the manual?

Keithley SourceMeter 2400 Series Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2400 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals