2400 Series SourceMeter
®
User’s Manual Sweep Operation 10-31
Call OutputCmd(intGPIB, “*RST”) 'Restore GPIB default
conditions.
Call OutputCmd(intGPIB, “:CALC2:CLIM:BCON END”)
Call OutputCmd(intGPIB, “:SENSE:FUNC:CONC OFF”) 'Turn off Concurrent
Functions.
Call OutputCmd(intGPIB, “TRIG:COUN 3”) 'Trigger count = #
sweep points.
Call OutputCmd(intGPIB, “:SOUR:FUNC MEM”) 'Source Memory sweep
Mode.
Call OutputCmd(intGPIB, “:SOUR:MEM:POIN 3”) 'Number of Memory
Points = 3.
Call OutputCmd(intGPIB, “:SOUR:MEM:STAR 1”) 'Start @ Memory Loca-
tion 1.
Call OutputCmd(intGPIB, “SOUR2:CLE:AUTO ON;:SOUR2:CLE:AUTO:DEL 1”)
'Set Digital I/O
auto-clear
Call OutputCmd(intGPIB, “:CALC2:cLIM:PASS:SOUR2 12” 'Set Digital Output
Pattern for Limit #2
Passing
Call OutputCmd(intGPIB, “OUTPUT ON”) 'Turn on Output
Call OutputCmd(intGPIB, “:INIT”) 'Trigger Sweep.
End Sub
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com