2400 Series SourceMeter
®
User’s Manual SCPI Command Reference 18-53
Sorting mode status bit values:
Result
1
Bit #: 21 20 19 9 8
Meas. Event
Status
2
Limit 1 and 4 pass and 2, 3, and
5-12 disabled
0 0 0 0 0 Bit 5 (LP)
Limit test 1 fail 0 0 0 0 1 Bit 0 (L1)
Limit test 2 pass 0 0 0 1 0 Bit 5 (LP)
Limit test 3 pass 0 0 0 1 1 Bit 4 (HL3)
Limit test 4 fail 0 0 1 0 1 Bit 10 (CC)
Limit test 5 pass 0 0 1 0 0 Bit 5 (LP)
Limit test 6 pass 0 0 1 1 0 Bit 5 (LP)
Limit test 7 pass 0 0 1 1 1 Bit 5 (LP)
Limit test 8 pass 0 1 0 0 0 Bit 5 (LP)
Limit test 9 pass 0 1 0 0 0 Bit 5 (LP)
Limit test 10 pass 0 1 0 1 0 Bit 5 (LP)
Limit test 11 pass 0 1 0 1 1 Bit 5 (LP)
Limit test 12 pass 0 1 1 0 0 Bit 5 (LP)
Limit 1 and 4 pass and 2, 3, and
5-12 fail
1 1 1 1 1 -
1. Limit 4 test available only with contact check option (Appendix F).
2. See Section 15, Measurement Event Register” and Figure 15-6 for details.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2400 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2400 Series and is the answer not in the manual?

Keithley SourceMeter 2400 Series Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2400 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals