Series 2600 System SourceMeters Reference Manual Range, Digits, Speed, Rel, and Filters 6-13
Return to Section 6 topics 2600S-901-01 Rev. A / May 2006
Filter configuration menu
Press CONFIG and then FILTER to display the filter configuration menu:
• AVERAGE-TYPE — Use this menu item to select filter type
(MOVING or REPEAT).
• AVERAGE-COUNT — Use this menu item to specify filter count
(1 to 100 readings).
Enabling filter
The filter is enabled by pressing the FILTER key. The FILT annunciator is on while
the filter is enabled. Pressing FILTER a second time disables filter.
Response time
The filter parameters have speed and accuracy trade-offs for the time needed to
display, store, or output a filtered reading. These affect the number of reading
conversions for speed versus accuracy and response to input signal changes.