12-96 Instrument Control Library Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006 Return to Section 12 topics
status.measurement.*
status.measurement.condition
status.measurement.enable
status.measurement.event
status.measurement.ntr
status.measurement.ptr
* = condition, enable, event, ntr or ptr
Attribute Measurement event status register set.
Usage Reads condition, enable, event, NTR and PTR registers:
measreg = status.measurement.condition
measreg = status.measurement.current_limit.enable
measreg = status.measurement.current_limit.event
measreg = status.measurement.current_limit.ntr
measreg = status.measurement.current_limit.ptr
Writes to enable, NTR and PTR registers:
status.measurement.enable = measreg
status.measurement.ntr = measreg
status.measurement.ptr = measreg
Set measreg to one of the following values:
0 Clears all bits.
status.measurement.VOLTAGE_LIMIT Sets VLMT bit (B0).
status.measurement.VLMT Sets VLMT bit (B0).
status.measurement.CURRENT_LIMIT Sets ILMT bit (B1).
status.measurement.ILMT Sets ILMT bit (B1).
status.measurement.READING_OVERFLOW Sets ROF bit (B7).
status.measurement.ROF Sets ROF bit (B7).
status.measurement.BUFFER_AVAILABLE Sets BAV bit (B8).
status.measurement.BAV Sets BAV bit (B8).
status.measurement.OUTPUT_ENABLE Sets OE bit (B11).
status.measurement.OE Sets OE bit (B11).
status.measurement.INSTRUMENT_SUMMARY Sets INST bit (B13).
status.measurement.INST Sets INST bit (B13).
measreg can also be set to the decimal weight of the bit to be set. Examples:
To set bit B0 (VLMT), set measreg to 1 (2
0
).
To set bit B1 (ILMT), set measreg to 2 (2
1
).
To set bit B8 (BAV), set measreg to 256 (2
8
).
To set more than one bit of the register, set measreg to the sum of their decimal
weights. For example, to set bits B0 and B8, set