Config InitiatePulseTest InitiatePulseTestDual
Source autorange (I&V) off. Output on. Output on.
Measure autorange (I&V) off. Enough free space in buffer. Enough free space in buffer.
Measure NPLC < ton. Buffer appendmode on when
pulse train is >1 point.
Buffer appendmode on when pulse
train is >1 point.
Measure autozero OFF or
ONCE.
Safety interlock engaged when
using the 200V range.
Safety interlock engaged when
using the 200V range.
Different unique SMUs for each
tag.
Same NPLC setting for each tag.
Same toff for each tag.
tag1 ton > tag2 ton by at least
41usec.
13-14 Factory Scripts Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006 Return to Section 13 topics