Series 2600 System SourceMeters Reference Manual Factory Scripts 13-33
Return to Section 13 topics 2600S-901-01 Rev. A / May 2006
Return
Values:
f: A Boolean flag. This flag will be true when the pulse was
successfully configured, false when errors were encountered.
msg: A string message. If the f flag is false, msg will contain an error
message. Otherwise, msg will contain a string indicating successful
configuration.
Output
Data:
None
Example(s):
smua.reset()
smua.source.rangev = 5
smua.source.rangei = 1
smua.source.levelv = 0
smua.measure.rangev = 5
smua.measure.rangei = 1
smua.measure.nplc = 0.01
smua.measure.autozero = smua.AUTOZERO_ONCE
smua.nvbuffer1.clear()
smua.nvbuffer1.appendmode = 1
smua.source.output = smua.OUTPUT_ON
smub.reset()
smub.source.func = smub.OUTPUT_DCAMPS
smub.source.rangei = 1
smub.source.rangev = 5
smub.source.leveli = 0
smub.measure.rangei = 1
smub.measure.rangev = 5
smub.measure.nplc = 0.01
smub.measure.autozero = smub.AUTOZERO_ONCE
smub.nvbuffer1.clear()
smub.nvbuffer1.appendmode = 1
smub.source.output = smub.OUTPUT_ON