Table 15-1
Recommended verification equipment
Description
Manufacturer/
Model
Accuracy
Digital Multimeter Agilent 3458A DC Voltage
1
(2601/2602)
DC Voltage
1
(2611/2612)
DC current
1
90mV:
0.9V:
5.4V:
36V:
190mV:
1.8V:
18V:
180V:
90nA:
0.9µA:
9µA:
90µA:
0.9mA:
9mA:
90mA:
0.9A:
±8ppm
±5ppm
±4ppm
±6ppm
±5ppm
±4ppm
±6ppm
±6ppm
±430ppm
±45ppm
±25ppm
±23ppm
±20ppm
±20ppm
±35ppm
±110ppm
0.5Ω, 250W, 0.1%
Precision Resistor
Isotek
RUG-Z-R500-0.1-TK3
Resistance
2
0.5Ω: ±125ppm
1. 90-day specifications show full-range accuracy of recommended model used for specified
measurement point.
2. Resistor used to test Model 2601/2602 3A range and Model 2611/2612 1.5A range only should be
characterized to uncertainty shown using resistance function of digital multimeter before use.
15-4 Performance Verification Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006 Return to Section 15 topics
Verification limits
The verification limits stated in this section have been calculated using only the
Series 2600 one-year accuracy specifications, and they do not include test
equipment uncertainty. If a particular measurement falls outside the allowable
range, recalculate new limits based both on the Series 2600 specifications and
corresponding test equipment specifications.

Table of Contents

Other manuals for Keithley SourceMeter 2612

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2612 and is the answer not in the manual?

Keithley SourceMeter 2612 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2612
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals