7.2 Adjustment bulk weighing balance, not preconfigured by KERN
Operation Display
Switch-on balance and during the self test press
.
Use the number keys to enter password:
Either
Standard password „0000“
or
personal password, input see chapter 13
Confirm input with .
Select bulk material balance using .
The appeared [T] displays the current balance.
Acknowledge with .
If necessary, at balance zero display using
select the weighing unit [kg or lb], which shall be
used for adjustment. The appeared [T] displays the
current weighing unit.
Acknowledge with .
Ensure that there are no objects on the weighing
plate. After successful stop check press the but-
ton.
Enter value of the necessary adjustment weight
using numeric keyboard and acknowledge with .
In order to achieve high-quality weighing results in
the sense of the measuring technology, it is recom-
mended to select the nominal value as high as pos-
sible. We recommend 80 % max.
Place the required adjusting weight carefully in the
centre of the weighing plate and press
to con-
firm.
After the adjustment the balance will carry out a self-
test. Remove adjusting weight during self test,
balance will return into weighing mode automatically.
In case of an adjustment error or incorrect adjusting
weight the display will show an error message
(
), repeat adjustment process.
CFS/CCS-BA-e-1013 19