17.1 Error messages
Error message Description Possible causes
Zeroing range exceeded
due to switching-on balance
or pressing (normally
4% max)
Object on the weighing
plate
Overload when zeroing
Improper adjustment
Damaged weighing cell
Damaged electronics
Keyboard error
Improper operation of
the balance
Value outside the A/D
changer range
Weighing plate not in-
stalled
Damaged weighing cell
Damaged electronics
Adjustment error
Improper adjustment
Should other error messages occur, switch balance off and then on again. If the
error message remains inform manufacturer.