EasyManuals Logo
Home>KERN>Scales>CFS

KERN CFS Operating Manual

KERN CFS
46 pages
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #46 background imageLoading...
Page #46 background image
17.1 Error messages
Error message Description Possible causes
Zeroing range exceeded
due to switching-on balance
or pressing (normally
4% max)
Object on the weighing
plate
Overload when zeroing
Improper adjustment
Damaged weighing cell
Damaged electronics
Keyboard error
Improper operation of
the balance
Value outside the A/D
changer range
Weighing plate not in-
stalled
Damaged weighing cell
Damaged electronics
Adjustment error
Improper adjustment
Should other error messages occur, switch balance off and then on again. If the
error message remains inform manufacturer.
46 CFS/CCS-BA-e-1013

Table of Contents

Other manuals for KERN CFS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN CFS and is the answer not in the manual?

KERN CFS Specifications

General IconGeneral
BrandKERN
ModelCFS
CategoryScales
LanguageEnglish

Related product manuals