CIB-BA-e-2010 21
7.5 Store/call up reference weight
50 memory locations (+ 20 quick memory locations) are available.
7.5.1 Save via numeric keyboard
Unload the weighing plate and enter the
reference weight via the numeric
keyboard. Then press longtime the
button .
(example)
Then press longtime the button .
„S 00“ is displayed flashing.
Enter the desired memory location via
the numeric keyboard and confirm by
.
The memory location has now been
allocated by the reference weight, the
balance is in weighing mode.