English
15.3.3 Automatic data output < auto >
Data output happens automatically without having to press the PRINT button as soon
as the corresponding output condition has been met, dependent on the setting in the
menu.
Enable function and set the output condition:
In Setup menu invoke the menu setting < print > < prmode> <
trig > and confirm with button.
For an automatic data output select the menu setting < auto > using the navi-
gation keys and confirm by the button.
Use the navigation keys to select the setting < on > and confirm on but-
ton. < zRange> is displayed.
Acknowledge by button and set the required output condition with the navi-
gation keys .
Acknowledge by button.
To exit the menu press the navigation key repeatedly.
Place goods to be weighed on balance:
If required, place empty container on scale and tare.
Place weighed goods and wait until the stability display ( ) appears.
The weighing value is issued automatically.
15.3.4 Continuous data output < cont >
Enable function and set the output interval:
In Setup menu invoke the menu setting < print > < prmode> <
trig > and confirm with button.
For a continuous data output select the menu setting < cont > using the navi-
gation keys and confirm on button.
Use the navigation keys to select the setting < on> and confirm on but-
ton.
< speed> is displayed.
Acknowledge with the button and set the required time interval with the navi-
gation keys (numerical input see chap. 3.2.2)
Set the required output condition <Zero> & <Stable>.
To exit the menu press the navigation key repeatedly.
Place goods to be weighed on balance
If required, place empty container on scale and tare.
Place goods to be weighed.
The weighing values are issued according to the defined interval.

Table of Contents