20 TPCB-A-BA-e-2314
Place the adjustment weight and confirm with button,
< wait> followed by < reMvld> will be displayed.
Once < reMvld> is displayed, remove the adjustment
weight.
After successful adjustment the balance automatically re-
turns to weighing mode.
In case of an adjustment error (e.g. objects on the weighing
plate) the display will show the error message < wrong
>. Switch off balance and repeat the adjustment process.
7.9.2 External adjustment with user-defined adjustment weight < caleud >
Press and hold the TARE and ON/OFF buttons simultane-
ously to enter the setup menu.
Wait until the first menu item < Cal > is displayed.
Confirm by button, < CalExt> will be displayed.
Use the navigation keys to select < caleud>.
Acknowledge by button. The numeric input window for
the weight value of the adjustment weight appears. The ac-
tive digit is flashing.
Provide adjustment weight.
Enter weight value, numerical input see chap. 3.2.2