EasyManuals Logo

KERN PCB User Manual

KERN PCB
66 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #36 background imageLoading...
Page #36 background image
36 TPCB-A-BA-e-2314
10.4.2 Weighing with multiplication factor via the application unit <FFA>
Here you determine with which factor the weighing result (in gram) will be multiplied.
By that way, e.g. a known error factor in the weight determination can be immediately
taken into account.
Select menu setting < unit> and confirm on button.
Use the navigation keys to select the setting < FFA >
and confirm on button.
Enter multiplication factor, numerical input
see chap. 3.2.2, the active digit flashes.
10.4.3 Percent weighing by application unit <%>
The application unit <%> allows to check the weight of a sample in percent, based on
a reference weight.
Select menu setting < unit>.
Place a reference weight which corresponds to 100 %
at the upper display
margin
at the upper display
margin
Acknowledge by button.
Use the navigation keys to select the setting < % > and
confirm on button.
Confirm the flashing weight value of the reference weight
using button.
From now on the weight of the sample will be shown in
percent based on the reference weight

Table of Contents

Other manuals for KERN PCB

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN PCB and is the answer not in the manual?

KERN PCB Specifications

General IconGeneral
CategoryPrecision balance
CalibrationExternal
DisplayLCD
Weighing unitsg, oz, lb
Housing materialPlastic
MaterialStainless steel
InterfaceRS232

Related product manuals