26 Keysight E5061B Network Analyzers
Troubleshooting
No Display Troubleshooting
2-
Step 6. Check around the backlight
Check inverter board and a cable between the inverter board and A52 Front
Panel I/F board. Also check the cables between the LCD and A52 Front Panel
I/F board. If the cables are normal, check the LCD.