EasyManua.ls Logo

Keysight Technologies E5061B

Keysight Technologies E5061B
121 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
92 Keysight E5061B Network Analyzers
Replacement Procedure
Power Switch Assembly Replacement
4-
Power Switch Assembly Replacement
Tools Required
torque screwdriver, T8, T15 and T20
torque screwdriver, T10 (set to 9 in-lb)
•flat edge screwdriver
Removal Procedure
Refer to Figure 4-21 for this procedure.
Step 1. Remove the Outer Cover as described in “Outer Cover Replacement” on
page 55.
Step 2. Remove the Front Panel as described in “Front Panel Replacement” on
page 56.
Step 3. Remove the Touch Screen Control Board as described in “Touch Screen
Control Board Replacement” on page 89.
Step 4. Remove the front panel rear side cover as described in “Inverter Board
Replacement” on page 90.
Step 5. Disconnect the connector (item 1) of cable from the A52 Front Panel Interface
Board.
Step 6. Remove the two TORX T10 screws (item 2) fastening the Power Switch
Assembly.
Step 7. Remove the parts (item 3) of the Power Switch Assembly.
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
Set a torque to 9 in-lb for torque driver when fasten the TORX T10 screws
(item 2).