56 Keysight E5061B Network Analyzers
Replacement Procedure
Front Panel Replacement
4-
Front Panel Replacement
Tools required
torque screwdriver T20
torque screwdriver T15 (set to 14 in-lb)
•flat edge screwdriver
Removal Procedure
Refer to Figure 4-2 for this procedure.
Step 1. Remove the Outer Cover as described in
“Outer Cover Replacement” on page 55.
Step 2. Remove the top trim (item 1) using a flat edge driver.
Step 3. Remove the seven TORX T15 screws (item 2 ) fastening front panel.
Step 4. Gradually push the front panel assembly towards the outside.
Step 5. Release the flat cable (item 3) from the cable clamp (item 4).
Step 6. Disconnect the flat cable (item 3) from the front panel.
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
Set a torque to 14 in-lb for torque driver when fasten the TORX T15 screws
(item 2).