Optionally, set a current limit value of 5 A while in CV mode:
CURR:LIM 5, (@1)
To select a lower current, power, or resistance range, program a value that falls within the range:
CURR:RANG 5, (@1)
RES:RANG 50, (@1)
POW:RANG 5, (@1)
To short the input terminals:
INP:SHOR ON, (@1:2)
To program a low-voltage limit (under-voltage inhibit) of 2 V, and enable the inhibit mode:
VOLT:INH:VON 2, (@1)
VOLT:INH:VON:MODE LIVE, (@1)
To set the current slew rate to 2 Amps/second:
CURR:SLEW 5, (@1)
To set the negative current slew, turn coupling (tracking) off. Then set the negative current slew:
CURR:SLEW:COUP OFF, (@1)
CURR:SLEW:NEG 3, (@1)
To set the sense terminals to remote sensing:
VOLT:SENS:SOUR EXT, (@1)
To query the setting of the sense terminals:
VOLT:SENS:SOUR? (@1)
Keysight EL30000 Series User's Guide 77