Decreases the attenuation value for the attenuation control unit. To decrease
attenuation,
<stepsize> value needs to be entered. If no <stepsize> value is stated, it
sets the value to the current step size or 1 dB initial step size (when unit power is on).
[INPut:]ATTenuation:DECRement [<stepsize>]
Examples
To decrease the attenuation level by a step size of 2 dB:
:DECR 2
Error Messages
If
<stepsize> is beyond the range specified, a return error value will be clipped to the
minimum or maximum attenuation value. You may encounter the following two error
messages if the value is not within the specified range:
- + 12, “Value clipped to Max”
- + 11, “Value clipped to Min”
Related Commands
[INPut:]ATTenuation/INPut[:ATTenuation]
[INPut:]ATTenuation:INCRement/INPut[:ATTenuation]:INCRement
[INPut:]ATTenuation:STEPsize/INPut[:ATTenuation]:STEPsize
[INPut:]FREQuency?
During auto correction mode, it is recommended that you enter both attenuation and frequency
values. Both values are intercorrelated and important for selecting the correct attenuation level.
Query? No
Set? Yes
Set Parameters
Numeric representation format: Integer (for
example, 8)
Minimum n/a
Maximum n/a
Units dB