Remote Operations 6
Keysight J7201A/B/C Operating and Service Manual 75
*RST
Resets the attenuation control unit to the following factory-defined states:
- :ATT 121 (For J7201A/B)
- :ATT 101 (For J7201C)
*RST
Examples
To reset the instrument:
*RST
Related Commands
*CLS
*IDN?
*OPC?
*TST?
*WAI
*SRE
Enables the bits in the status byte enable register. The selected enabled bits are
summarized in the “Master Status Summary” (MSS) bit (bit 6) of the Status Byte register.
If any of the selected bit condition changes from 0 to 1, a service request is generated.
Table below shows the contents of this register.
The query returns the current contents of the status byte enable register. The format of
the return is in the range of 0 to 255.
Table 6-3 Bit configuration of Status Byte Enable register
Query? No
Set? Yes
Set Parameters n/a
Minimum n/a
Maximum n/a
Units None
Bit Position 765432-0
Bit Name OPER
MSS
RQS
ESB MAV QUES n.u.