6 Remote Operations
60 Keysight J7201A/B/C Operating and Service Manual
Examples
To open all channels:
ROUTe:OPEn:ALL
Related Commands
ROUTe:CLOSe
ROUTe:OPEn
ROUTe:CLOSe:ALL
DIAGnostic:RELay:CYCLes?
Checks number of cycles for individual section. The first command checks number of
cycles for individual section. The second command shows number of cycles for all
sections. The following table lists numbers (1 - 8) representing each section of the
attenuators.
DIAGnostic:RELay:CYCLes? <channel_list>
Maximum n/a
Units None
ROUTe:CLOSe, ROUTe:OPEn, ROUTe:CLOSe:ALL and ROUTe:OPEn:ALL SCPI commands are
not supported in Auto correction mode.
Table 6-1 Sections of attenuators
Attenuator A Attenuator B
Sec 1 1 Sec 1 5
Sec 2 2 Sec 2 6
Sec 3 3 Sec 3 7
Sec 4 4 Sec 4 8
The first command above returns number of cycles for Sec 1 for attenuator A and Sec 4 for
attenuator B. Number of cycles for all sections will be displayed when second command is send.
Query? Yes, query only
Set? No
Return Parameters
Numeric representation format: Integer (for
example, 8)
Minimum 101