6 Remote Operations
54 Keysight J7201A/B/C Operating and Service Manual
[INPut:]ATTenuation:INCRement/INPut[:ATTenuation]:INCRement
Increases the attenuation value for the attenuation control unit. To increase attenuation,
<stepsize> value needs to be entered. If no <stepsize> value is stated, it sets the value
to the current step size or 1 dB initial step size (when unit power is on).
[INPut:]ATTenuation:INCRement [<stepsize>]
Examples
To increase the attenuation level by a step size of 4 dB:
:INCR 4
Error Messages
If
<stepsize> is beyond the range specified, a return error value will be clipped to the
minimum or maximum attenuation value. You may encounter the following two error
messages if the value is not within the specified range:
- + 12, “Value clipped to Max”
- + 11, “Value clipped to Min”
Related Commands
[INPut:]ATTenuation/INPut[:ATTenuation]
[INPut:]ATTenuation:DECRement/INPut[:ATTenuation]:DECRement
[INPut:]ATTenuation:STEPsize/INPut[:ATTenuation]:STEPsize
[INPut:]FREQuency?
[INPut:]ATTenuation:STEPsize/INPut[:ATTenuation]:STEPsize
Sets the step size for the attenuation control unit. The initial step size is set to 1 dB when
the unit is powered on. If step size is set to some other value, for example, 10 dB,
increase or decrease attenuation level is following 10 dB step size.
[INPut:]ATTenuation:STEPsize <numeric>
Query? No
Set? Yes
Set Parameters
Numeric representation format: Integer (for
example, 8)
Minimum n/a
Maximum n/a
Units dB
Query? Yes
Set? Yes
Set/Return Parameters
Numeric representation format: Integer (for
example, 8)
Minimum n/a