6 Remote Operations
58 Keysight J7201A/B/C Operating and Service Manual
a colon separated range of channels as in: 101:107; a combination of channels and
ranges as in: 102,103,104:107.
ROUTe:CLOSe <channel_list>
Examples
To close the channel number 106:
ROUTe:CLOSe (@106)
To query the closed list of channels:
ROUTe:CLOSe? <channel_list>
Related Commands
ROUTe:OPEn
ROUTe:CLOSe:ALL
ROUTe:OPEn:ALL
ROUTe:OPEn
Opens the switching path(s). The first parameter specifies the Channels in the form of
‘bcc’ where b = bank number (0-n, range model dependent) and cc = channel number
(00-nn, range model dependent). For example: 103 or 104 or 108. A list of channels may
be specified in several forms: a comma separated list of channels as in: 101,103,105,108;
a colon separated range of channels as in: 101:107; a combination of channels and
ranges as in: 102,103,104:107.
ROUTe:OPEn <channel_list>
Query? Yes
Set? Yes
Set/Return Parameters
Numeric representation format: Integer (for
example, 8)
Minimum 101
Maximum 108
Units None
Query? Yes
Set? Yes
Set/Return Parameters
Numeric representation format: Integer (for
example, 8)
Minimum 101
Maximum 108
Units None