Remote Operations 6
Keysight J7201A/B/C Operating and Service Manual 57
Error Messages
If <numeric> is beyond the range specified, a return error value will clipped to minimum
or maximum attenuation value. You may also encounter 2 error messages if value is not
within range specified. If relative is not turn on, it will return error +13.
- + 12, “Value clipped to Max”
- + 11, “Value clipped to Min”
- + 13, "Not in relative mode”
Related Commands
[INPut:]ATTenuation:RELative[:STATe]/INPut[:ATTenuation]:RELative[:STATe]
[INPut:]ATTenuation:REFerence/INPut[:ATTenuation]:REFerence
[INPut:]FREQuency?
[INPut:]FREQuency
Set the instrument frequency. The return value is in MHz.
[INPut:]FREQuency <float64>
Examples
To set the frequency to 25.5 GHz:
FREQ 25.5GHz
To query the set frequency:
[INPut:]FREQuency?
Related Commands
[INPut:][ATTenuation:]RELative[:STATe]
[INPut:][ATTenuation:]REFerence
ROUTe:CLOSe
Closes the switching path(s). The first parameter specifies the Channels in the form of
‘bcc’ where b = bank number (0-n, range model dependent) and cc = channel number
(00-nn, range model dependent). For example: 103 or 104 or 108. A list of channels may
be specified in several forms: a comma separated list of channels as in: 101,103,105,108;
Query? Yes
Set? Yes
Set/Return Parameters
Numeric representation format: This
parameter is a number that can be expressed
as an integer (for example, 8) or a real number
(for example, 23.6)
Minimum 0 Hz
Maximum
6 GHz (J7201A)
18 GHz (J7201B)
26.5 GHz (J7201C)
Units Hz