6 Remote Operations
56 Keysight J7201A/B/C Operating and Service Manual
[INPut:]ATTenuation:REFerence/INPut[:ATTenuation]:REFerence
Sets the reference attenuation for the attenuation control unit in relative attenuation
mode. If integer value is not given, it automatically references to current attenuation as
reference level.
[INPut:]ATTenuation:REFerence [<numeric>]
Examples
To set the reference attenuation to 12:
:REF 12
Related Commands
[INPut:]ATTenuation:RELative[:STATe]/INPut[:ATTenuation]:RELative[:STATe]
[INPut:]ATTenuation:RELative:LEVel/INPut[:ATTenuation]:RELative:LEVel
[INPut:]FREQuency?
[INPut:]ATTenuation:RELative:LEVel/INPut[:ATTenuation]:RELative:LEVel
Sets user desired relative attenuation. This relative attenuation level will be referenced to
reference attenuation level set earlier.
[INPut:]ATTenuation:RELative:LEVel <numeric>
Examples
To set the relative attenuation to 28:
:REL:LEV 28
To query the set relative attenuation:
[INPut:]ATTenuation:RELative:LEVel?
Query? No
Set? Yes
Set Parameters
Numeric representation format: Integer (for
example, 8)
Minimum n/a
Maximum n/a
Units n/a
Query? Yes
Set? Yes
Set/Return Parameters
Numeric representation format: Integer (for
example, 8)
Minimum n/a
Maximum n/a
Units None